Operator’s Guides — Medical Device Cybersecurity · Module 2 of 17

Law, Guidance, or Good Practice? How to Tell Which FDA Cybersecurity Requirements Bind You

Every page of the FDA's cybersecurity guidance is stamped Contains Nonbinding Recommendations. Underneath the stamp sit a binding quality regulation, four rules of federal law for cyber devices, and a shelf of optional standards.

All Medical Device Cybersecurity modules
Module 02 9:28 video + article

Watch this module

Module 2 of the video course. The article below covers the same ground in written form, so you can watch, read, or both.

Somebody in your planning meeting says the FDA requires it. The item on the table might be a penetration test, a software bill of materials, or a risk process built on a standard called AAMI TIR57. If your device counts as what the law calls a cyber device, one of those three is federal law, one is the FDA's recommendation, and one is a standard the FDA lists as an option.

The FDA's premarket cybersecurity guidance prints three words at the top of every page: Contains Nonbinding Recommendations. The stamp is accurate. It doesn't make the guidance safe to ignore, because the guidance sits on rules that do bind.

What the stamp covers

The guidance explains the stamp on its own title page. The document, it says, doesn't create rights for anyone, and it doesn't bind the FDA or the public. When it says a manufacturer should do something, it is making a recommendation. The same page allows a different approach, as long as that approach satisfies the statutes and regulations that apply. So the guidance sits on top of the statutes and regulations, and it describes one way to meet them.

The regulation under the guidance

One of the regulations under the guidance covers device makers directly. The FDA publishes a fact sheet on common myths about device cybersecurity, and one myth it answers is that cybersecurity is optional. The agency's answer is the quality system regulation, which requires manufacturers to address risks, cybersecurity risk included. The guidance offers recommendations for meeting that requirement. In February 2026 the FDA rebuilt the regulation on ISO 13485, the international standard for quality management, and brought the standard in by reference. That makes ISO 13485 a standard a device maker is bound by.

So there are three tiers of obligation: the law and its regulations, ISO 13485 now among them, which bind you; the FDA's guidance, which you can argue with; and the optional standards, which you pick.

This is part two of The Product Manager's Guide to Cybersecurity in Medical Devices, and it stands on its own. The series follows one made-up device through all three tiers: the Tidewell patch, a heart monitor from a fictional company called Northwater Medical, which syncs to the patient's phone over Bluetooth. Tidewell's security risk work already sits in two tiers. The quality system regulation requires that work, and the guidance describes how to do it.

Four rules of federal law

Congress added a cybersecurity section to the law at the end of 2022. On December 29, a year-end spending bill put section 524B into the Federal Food, Drug, and Cosmetic Act. Section 524B covers what it calls cyber devices, and it applies when a company submits one for the FDA's review. For now, picture a device with software that can connect to the internet.

For a cyber device, the law lists four things the company shall do:

1. Send the FDA a plan for finding and fixing vulnerabilities after launch, with a way for outside researchers to report what they find. 2. Keep processes that give a reasonable assurance the device and its related systems are secure, patches included. Vulnerabilities the company judges unacceptable get fixed on a regular schedule it can justify. Critical ones get fixed as soon as possible, outside that schedule. 3. Provide a software bill of materials, a list of the commercial, open-source and off-the-shelf software inside the device. 4. Meet any other requirement the FDA sets through a formal regulation, if the agency chooses to write one.

Failing the second rule, the one with the patches in it, is a prohibited act under the same statute. (I'm not a lawyer, and nothing here is legal advice.) Put the four rules beside the stamp on the guidance and compare the verbs: the law says shall, and the guidance says should.

The statute itself is short. It asks for a plan, processes and a bill of materials, and it says very little about what goes inside any of them. The FDA fills that in with Section VII of the premarket guidance, which lays out the information the agency expects to see for each rule. Now and then the guidance says must, when it is repeating the law: manufacturers must provide documentation to comply with section 524B, for example. That obligation belongs to the statute, and the advice that follows it goes back to should.

Two dates and a screening hold

Section 524B came with two dates. It applied to submissions sent from March 29, 2023. A temporary FDA policy about refusing to accept cyber device submissions ran until October 1 of that year, and after October 1 the FDA said it expected companies to have the required information ready.

October 1, 2023 was also the day electronic filing through a template called eSTAR became mandatory for most 510(k) submissions. The template has a cybersecurity section. If your answers there are inaccurate or incomplete, the FDA says the submission goes on a technical screening hold, before the scientific review begins.

Arguing with the guidance

A document stamped nonbinding still describes what your filing has to show under a binding law. You can argue with that advice, and the place to do it is with the FDA, before you file. The title page tells you to discuss a different approach with the agency staff it lists, and the guidance recommends a Q-Submission, a formal request for the FDA's feedback before you file, to talk through cybersecurity design early.

The way a pre-submission meeting was explained to me, you present your plan, the FDA reacts to it, and sometimes you have to read its answer between the lines. A regulatory attorney also told me that a yes in that meeting doesn't bind the team that reviews your submission later. Take a yes as a good sign, and still build your case for a reviewer who wasn't in the room.

Raise a departure from the guidance early, in writing and in conversation with the FDA, and that takes time on your schedule. Where a process you already run does the job, the guidance says the FDA isn't asking you to replace it.

The standards you pick

The third tier is the standards. For a secure development process, the guidance lists frameworks to consider, starting with the industry's Joint Security Plan, known as JSP2, and it says the list isn't complete. For security risk, it points to AAMI TIR57. These are tools for doing the work the guidance describes, and some of them cost money just to read. Section 524B names none of them.

Mistaking an optional standard for law funds work nobody required. The opposite mistake, treating the guidance as optional, leaves you answering to the quality system regulation anyway, and it can put your filing on hold.

Tidewell, tier by tier

Assuming Tidewell is a cyber device, Northwater's obligations sort into the three tiers:

Tidewell's obligations, by tier 1. The law and its regulations, which bind Northwater: the quality system regulation; Part 806, a rule that gives Northwater ten working days to report a fix made to reduce a risk to health; and section 524B's four rules, the software bill of materials among them. 2. The FDA's guidance, which Northwater can argue with: what Section VII expects inside each rule, recommendations such as security testing, and a threat model (a written study of how someone could attack the device). 3. The standards, which Northwater picks: JSP2 is one candidate, since the guidance names it and the plan is publicly available.

Northwater funds the second tier's work, or it books a meeting early to propose something else. Sorted this way, the three items from the planning meeting land in three places: the bill of materials is law, the penetration test is the FDA's recommendation, and TIR57 is a choice. Whether Tidewell is a cyber device at all is the subject of episode three, which takes it through the law's three-part definition.

The next time someone says the FDA requires it, ask which tier it's in. Then find the page it came from, and check the top of that page for the stamp.

The video version of this episode is on YouTube at https://youtu.be/x5PmM1AnQkI.

Get the next guide

ISO 13485, IEC 60601, and the GHG Protocol are in the queue. Leave your email and each new guide lands in your inbox the day it ships. Nothing else, no drip sequence.

Done. You’ll get the next guide when it ships.

The course is free. If you want help running product on a connected device, that’s the work I do.

Work With Dave

Prefer a smaller first step? Book a $500 one-hour working session →

Dave Saunders

Dave Saunders is the founder of Base Reality Group and a Fractional CPO for product companies. He was a founder and operator at Galen Robotics, where the surgical-robotics platform earned FDA De Novo authorization in 2023, and he managed a 35-patent portfolio licensed from Johns Hopkins. He wrote Founders Who Finish and publishes The Build. More about Dave →