Operator’s Guides — Medical Device Cybersecurity · Module 3 of 17

Is Your Medical Device a Cyber Device? What Section 524B Says, What the FDA Reads Into It, and When It Starts

A device that only pairs with a phone over Bluetooth never goes online, and on the FDA's reading of one word in the law it can still be a cyber device. The duties that follow wait for the next filing.

All Medical Device Cybersecurity modules
Module 03 10:18 video + article

Watch this module

Module 3 of the video course. The article below covers the same ground in written form, so you can watch, read, or both.

Sometime this week you probably paired something with your phone over Bluetooth: earbuds, a watch, the car. It didn't feel like going online, and in the plain sense it wasn't. If the thing you paired were a medical device, though, the FDA would count that radio as the ability to connect to the internet.

This is episode three of The Product Manager's Guide to Cybersecurity in Medical Devices, and it stands on its own. The series follows a made-up heart monitor, the Tidewell patch, from a fictional company called Northwater Medical. Tidewell's spec sheet describes its connection in one short line: it syncs to the phone over Bluetooth.

Blue Goat Cyber, a security firm, says a lot of its clients argue their device isn't a cyber device, often because a USB port is only used for maintenance. Whether they're right depends on one definition in federal law, which has three parts. The middle one turns on the word ability.

Three parts, joined by "and"

Congress wrote the definition into section 524B of the Federal Food, Drug, and Cosmetic Act. A cyber device is a device that includes software, has the ability to connect to the internet, and contains something that could be vulnerable to cybersecurity threats. The three parts are joined by and, so a device has to meet all of them.

When a company files for a device that meets all three, the law's four rules apply. The company owes the FDA a plan for handling vulnerabilities after launch, processes that keep the device secure (patches included), and a software bill of materials, which is a list of the software inside. The fourth rule covers anything else the FDA chooses to set by formal regulation.

Part one: software, as the FDA reads it

The statute's first part covers software that the sponsor validated, installed, or authorized, either as a device or in a device. The sponsor is the company bringing the device to the FDA. That's as far as the law goes. The FDA's premarket cybersecurity guidance adds a reading: software includes firmware, the code built into the device itself, and programmable logic, the instructions stored on a chip.

On that reading, a team that ships nothing but firmware still meets part one. Tidewell's firmware runs its Bluetooth radio, and Northwater put it there, so Tidewell meets part one too.

Part two: the word "ability"

The second part says "the ability to connect to the internet," and the law never defines the phrase. Episode two of the series sorted a company's obligations into the law, the FDA's guidance, and optional standards. The word ability is law. What it covers is the FDA's reading.

The FDA reads it broadly. The guidance says a device counts if it can reach the internet through any means, intentionally or unintentionally. The agency's reasoning is that a device with the ability can end up connected. The guidance then lists examples and says the list isn't complete:

The maintenance port from Blue Goat's clients gets a footnote of its own. A device may need to be serviced over a USB connection, the FDA writes, and while that connection may be brief, the ability to connect is present. The footnote is silent on how hard the port is to reach. If someone on your team says a sealed port keeps you outside section 524B, ask them where the FDA says so.

Tidewell's spec sheet says the patch talks to the phone over Bluetooth, and the phone does the uploading. A team that says the patch never goes online is describing the patch correctly. The FDA reads the law as asking what the patch is able to do. Bluetooth is on the agency's list, so Tidewell meets part two.

That reading is guidance, so it can be argued with. Arguing means arguing with the agency that wrote it, and the time to do that is before you file. The FDA's own FAQ says a manufacturer unsure whether its device is a cyber device may contact the agency. (I'm not a lawyer, and nothing here is legal advice.)

Part three: "could be vulnerable"

The third part asks whether anything the sponsor validated, installed, or authorized in the device could be vulnerable to cybersecurity threats. The word could sets a low bar, and the third part is still a test in the law, so answer it in writing. For Tidewell the answer is the Bluetooth link and the firmware behind it.

With all three parts met, Tidewell is a cyber device on the FDA's reading. So far nothing has changed for Northwater, because the law's duties attach to a filing.

When the rules arrive

Section 524B applies to whoever submits certain applications to the FDA for a cyber device. The guidance names them, and the ones a product team is most likely to meet are the 510(k), the De Novo, and the PMA, the filings that ask to bring a device to market. The FDA says the list also includes smaller filings for changes, such as a special 510(k) or a PMA supplement. Northwater plans a 510(k) for Tidewell, so that's where the four rules come due.

A device already on the market works differently. Take a Bluetooth heart monitor cleared in 2021. The FDA's FAQ says the law's cybersecurity requirements don't apply to a submission sent before March 29, 2023, so the filing that cleared that monitor needed none of this. But if a cleared cyber device gets a change that requires premarket review, the FAQ says, the law applies to the new submission. For a device already on the market, then, find the date of its next filing. Your regulatory lead can tell you which changes need one.

The FDA sorts those changes by whether they touch security. Its examples of changes that may affect cybersecurity are new connectivity features, new authentication or encryption algorithms, and a changed software update process. Add a Wi-Fi radio to that 2021 monitor, and if the change needs a filing, the FDA points you to the full set of cybersecurity documents.

The guidance treats other changes, such as a new material or a new sterilization method, as unlikely to affect cybersecurity. For those it recommends a lighter package of three items: the plan for handling vulnerabilities, if none was ever filed; the software bill of materials; and a short description in place of full documentation of the security processes.

Suppose the monitor's maker changes its sterilization method, and the change needs a filing. If no plan for handling vulnerabilities was ever filed, the FDA says the company must provide one now, along with the software bill of materials. The short description covers whether any critical vulnerabilities could cause an unacceptable risk of harm to patients, and whether the company has fixed any since the last clearance. For a company that never filed a plan, a filing about sterilization is where its first cybersecurity plan comes due.

Exemptions, and failing a part

The law lets the FDA publish a list of exempt devices in the Federal Register. I searched the Federal Register on November 6, 2026, and didn't find one, so check it yourself before you count on it. Without a listing, the way out of the definition is to fail one of its three parts.

A device that fails one drops the law's four rules. An FDA fact sheet says the agency's quality regulations still require manufacturers to address risks, cybersecurity risk included, and the FDA says its cybersecurity guidance isn't limited to network-enabled devices.

Put the answer on one page

Whatever the answer for your device, put it on one page. The format below is mine, and the FDA didn't ask for it:

1. The software, firmware, and programmable logic you put in the device. 2. Every network, radio, and connector, including the ones only a service tech uses. 3. Which of those could be vulnerable, and why. 4. Your decision, with what comes from the law and what comes from the FDA's reading kept apart. 5. The next filing that brings all of this in, with its date and the name of whoever signs.

Northwater's page is short. The patch runs Northwater's firmware, and its Bluetooth link could be vulnerable. The decision is cyber device, on the FDA's reading of the word ability, and the trigger is the 510(k).

To start your own, pull the spec sheet for your device and find the line that says Bluetooth, or USB, or service port. Copy it onto line two.

Northwater's line five names the 510(k), and the space for whoever signs is still blank. Episode four is about who in your company owns this work, and what that person needs to do the job. The video version of this episode is here: https://youtu.be/C1kb3Hllbn4.

Get the next guide

ISO 13485, IEC 60601, and the GHG Protocol are in the queue. Leave your email and each new guide lands in your inbox the day it ships. Nothing else, no drip sequence.

Done. You’ll get the next guide when it ships.

The course is free. If you want help running product on a connected device, that’s the work I do.

Work With Dave

Prefer a smaller first step? Book a $500 one-hour working session →

Dave Saunders

Dave Saunders is the founder of Base Reality Group and a Fractional CPO for product companies. He was a founder and operator at Galen Robotics, where the surgical-robotics platform earned FDA De Novo authorization in 2023, and he managed a 35-patent portfolio licensed from Johns Hopkins. He wrote Founders Who Finish and publishes The Build. More about Dave →