In July 2015 the FDA told hospitals to stop using an infusion pump called the Symbiq and to move their patients to other pumps. The agency wasn't aware of a single patient who had been hurt by it. What it objected to was the pump's network connection, which was one of the features hospitals had bought it for.
If you run product on a medical device that connects to anything, the Symbiq's problem can become yours. This is the first part of The Product Manager's Guide to Cybersecurity in Medical Devices, and it stands on its own.
Why these devices are connected
An infusion pump pushes medication into a patient's IV at a rate the care team sets. It works from a drug library, the set of medication settings the pump uses, and those settings change. Hospira, which made the Symbiq, put the pump on the hospital's network so the drug library could be updated from one central system. The connection spared a hospital's staff from updating every pump by hand.
Pacemakers and defibrillators talk to the outside world by radio, and Kevin Fu explained why to a conference audience in 2016. Fu is a security researcher whose group at the University of Michigan studied these devices. Before wireless, changing an implant's settings meant pushing a needle in through the patient's armpit to turn a dial on the implant itself. The danger Fu stressed with the needle was infection, so the radio went in to protect the patient.
Open ports and a factory password
Each of those connections is also a way into the device. Fu said most of what researchers like him find is like breaking down an open door.
The Symbiq had openings like that, and the FDA listed them. A network port is a numbered channel a device listens on for connections, and two of the Symbiq's ports ran FTP and Telnet, which are old ways of reaching into a computer over a network. A third port still accepted the password the pump had shipped with.
Researchers had already been publishing on Hospira's pumps. In May 2015, after a researcher published the weaknesses of two of the company's other pumps, software code included, the FDA warned hospitals about those pumps and told them to close the same two ports.
A head of product sits in the room where choices like these get made. Services left running on open ports and a password unchanged from the factory are design-stage decisions, made long before any researcher goes looking.
A wrong dose
The FDA spelled out what could happen through the Symbiq's ports. An unauthorized user could take control of the pump and change the dosage it delivers, which, in the agency's words, "could lead to over- or under-infusion of critical patient therapies."
A wrong dose is what the FDA guards against. Patient data matters too, and a footnote in the agency's premarket cybersecurity guidance sets it apart: the loss of confidential health information is generally not considered a direct impact on safety and effectiveness.
The fix that moved the risk
For the Symbiq, the FDA strongly encouraged hospitals to move to other pumps, and until they could, it told them to take the pump off the network.
Taking the pump off the network switched off the feature the hospital had paid for, and the FDA said so in a box marked caution. Without the network, each pump's drug library had to be updated by hand, and manual updates, the agency wrote, "can be labor intensive and prone to entry error." The risk moved, from a hacker nobody had seen to the people typing medication settings into each pump.
The Symbiq was already on its way out. Hospira had announced in 2013 that it would retire the pump, and by the time of the FDA's notice it had stopped making and selling it, for unrelated reasons, so the advice to stop landed on a product that was leaving anyway. On a product you're still selling, the same finding means a patch, a recall or a replacement program, paid for out of your budget.
Hospitals can order a connection switched off too. I've talked with people who hold veto power over network access at some of the top hospitals in this country, and they've told me they won't allow a surgical robot a live internet connection during a procedure.
No known harm at St. Jude
A company can pay heavily for a weak connection when no patient is known to be hurt, and St. Jude Medical did, in public, starting in 2016.
St. Jude made pacemakers and defibrillators, and it made the Merlin@home transmitter, a box by the patient's bed that carried readings from the implant to the clinic. In August 2016 an investment firm called Muddy Waters published a report claiming the devices could be attacked. A security firm had done the research, and Muddy Waters disclosed that it was betting St. Jude's stock would fall.
The stock dropped more than 8% during the day and closed down nearly 5%. St. Jude was in the middle of being acquired by Abbott, and analysts began asking whether the deal was in danger. St. Jude called the claims "absolutely untrue" and sued Muddy Waters and the research firm.
Regulators examined the devices next. In January 2017 the FDA confirmed vulnerabilities in the Merlin@home transmitter that could let someone other than the patient's physician remotely access the implanted device. In August 2017 it approved an update to the pacemakers' firmware, the software built into the device itself. The update covered 465,000 pacemakers in the United States, and it couldn't be sent over the air, meaning wirelessly from a distance, so every patient needed an in-person visit.
The Department of Homeland Security's advisory on those pacemakers listed the weaknesses, including one that let someone within radio range keep sending wake-up commands until the battery ran down. The fix carried a clinical risk of its own. As a guide, the FDA published the malfunction rates from earlier firmware updates: in about 3 of every 100,000 of those updates, the device lost its function completely. Neither the FDA nor Abbott recommended replacing anyone's pacemaker over the vulnerability, and even applying the patch meant weighing one small risk against another.
From the Muddy Waters report to the firmware update, the FDA's statement on harm never changed: there were no known reports of patient harm. The company paid anyway, through its share price, through the lawsuit it filed, and through a fix that required a clinic visit for each patient.
Düsseldorf and the record on harm
The FDA keeps a list of its cybersecurity safety communications, and it says it isn't aware of any patient injuries or deaths in the cases on that list. In 2020, ransomware closed the emergency department at a hospital in Düsseldorf, and a critically ill patient was taken to another hospital about 20 miles away, where she died. Reporters called it the first death caused by a cyberattack.
Prosecutors in Cologne investigated, and the chief prosecutor concluded that the trip to the second hospital was "of no relevance to the final outcome." The patient's medical condition alone caused her death. That attack also hit the hospital's computer systems, with no medical device involved.
A job for the life of the device
Kevin Fu, in the same 2016 talk, said patients are far better off with these devices than without them. The job, then, is to keep the connection and budget for what it costs over the device's life.
The FDA's premarket guidance says no device is, or can be, completely secure. It also says that exploiting a known vulnerability, or a weak security control, should be treated as a reasonably foreseeable failure. Once you know about a weakness, the agency expects your design to deal with it.
The cost of a connection keeps coming for as long as the device is in use. New weaknesses turn up after launch, in code you shipped years ago, and researchers you've never met go looking for them. Each weakness they find comes back to the connection you specced, and someone at your company has to answer for it.
The FDA says that someone is the manufacturer, and its fact sheet on common myths says so twice. Premarket testing, it says, is the responsibility of the medical product manufacturer. A manufacturer that chooses to build on another company's software bears responsibility for that software's security. Nor can a manufacturer hand the device's security to the hospital's IT team. The FDA says hospitals may need to patch or change their networks to reduce risk, and it tells them to work closely with the device maker when they do.
Since 2023, the law
The core of the manufacturer's responsibility has been law since 2023. Congress added section 524B to the Federal Food, Drug, and Cosmetic Act, and it took effect that March for what the law calls cyber devices. I'm not a lawyer, and nothing here is legal advice.
The law's key requirement is that a manufacturer design, develop and maintain processes that give a reasonable assurance that the device and its related systems are cybersecure. The manufacturer must also make updates and patches available after launch. Failing that requirement is a prohibited act under the same law, and the word maintain carries the duty well past clearance.
The phrase related systems reaches further than the device. The FDA's guidance reads it to include things the manufacturer controls, and its examples include the servers that deliver software updates and the connections into a hospital's network. The duty follows the connection out of the box you ship and into the systems your company runs.
The law's definition of a cyber device starts with software and the ability to connect to the internet. Whether your device qualifies gets its own episode, episode three. Until then, treat any device that talks to anything as a candidate.
One device, followed the whole way
This series follows one made-up device from start to finish, so that nobody's real product gets picked apart. Northwater Medical is a company of about forty people, and its product is the Tidewell patch, a heart monitor a patient wears on the chest. Northwater and Tidewell are fictional.
Tidewell talks to the patient's phone over Bluetooth. The phone app sends the readings up to Northwater's cloud, and a cardiology clinic reviews them in a web dashboard on the hospital's network. The Bluetooth link does what Fu's radio did. The cloud does what the Merlin@home transmitter did for St. Jude. And the clinic dashboard sits on the same kind of hospital network the Symbiq had to come off.
On Tidewell's spec sheet, one line reads "syncs to the phone over Bluetooth." That line, and every connection on your own spec sheet, can go through five questions this week:
1. What does it talk to, and why does a patient need it? 2. Who chose its defaults (the passwords, the open ports, the encryption), and when? 3. What breaks for the clinician if the FDA or a customer has you switch it off tomorrow? 4. Who watches it after launch, and for how many years? 5. Does it make your device a cyber device?
If you only get to one this week, take the third. It's the question the Symbiq's hospitals had to answer in 2015, when unplugging the pump sent its drug library back to hand entry.
Episode two separates what the law requires from what the FDA only recommends, because mixing the two up leads either to overspending or to shipping something you can't defend. The video version of this episode is on YouTube at https://youtu.be/nVrSghz842Y.
Get the next guide
ISO 13485, IEC 60601, and the GHG Protocol are in the queue. Leave your email and each new guide lands in your inbox the day it ships. Nothing else, no drip sequence.
Done. You’ll get the next guide when it ships.
The course is free. If you want help running product on a connected device, that’s the work I do.
Work With DavePrefer a smaller first step? Book a $500 one-hour working session →
Dave Saunders is the founder of Base Reality Group and a Fractional CPO for product companies. He was a founder and operator at Galen Robotics, where the surgical-robotics platform earned FDA De Novo authorization in 2023, and he managed a 35-patent portfolio licensed from Johns Hopkins. He wrote Founders Who Finish and publishes The Build. More about Dave →